In vivo aging-induced surface roughness alterations of Invisalign<sup>®</sup> and 3D-printed aligners
Despina Koletsi, Nearchos Panayi, Christodoulos Laspos, Athanasios E. Athanasiou, Spiros Zinelis, Theodore Eliades
Abstract
Objective: To assess the surface roughness of in-house 3D-printed orthodontic aligners compared with Invisalign ® appliances, both retrieved as well as in the ‘as-received’ control status. Design: An in vitro study following intra-oral material aging. Setting and participants: Twelve clinically used Invisalign ® appliances and the same number of 3D-printed aligners, without involvement of attachments, were obtained from a respective number of patients. A similar number of ‘as-received’ aligners, of each material, were used as control (CON) groups. Method: Four groups of materials were examined: A = Invisalign ® CON; B = Invisalign ® used; C = 3D-printed CON; and D = 3D-printed used. Optical profilometry was employed to examine the following surface roughness parameters: amplitude parameters Sa, Sq and Sz and functional parameters Sc and Sv. Descriptive statistics and quantile regression modeling were conducted, and the level of statistical significance was set at α = 0.05. Results: Intra-oral exposure of 3D-printed aligners was significantly associated with increase in all tested parameters ( P < 0.001 at all occasions). Significant differences were detected in the retrieved 3D-printed aligners compared with Invisalign ® retrieved, with the exception of Sz. The respective effect sizes (median differences) were as follows: Sa: 169 nm, 95% confidence interval [CI] = 89–248, P < 0.001; Sq: 315 nm, 95% CI = 152–477, P < 0.001; Sc: 233 nm 3 /nm 2 , 95% CI = 131–335, P < 0.001; and Sv: 43 nm 3 /nm 2 , 95% CI = 17–68, P = 0.002. Conclusion: Within the limitations of this study, we concluded that surface roughness differences existed between 3D-printed aligners and Invisalign ® in the retrieved status, as well as between the control and retrieved 3D-printed groups.