Ceramic-on-ceramic <i>vs</i> ceramic-on-polyethylene, a comparative study with 10-year follow-up

Justin van Loon, Daniël Hoornenborg, Harm M van der Vis, Inger N. Sierevelt, Kim TM Opdam, Gino M. M. J. Kerkhoffs, Daniël Haverkamp

2021World Journal of Orthopedics19 citationsDOIOpen Access PDF

Abstract

BACKGROUND: In press-fit total hip arthroplasty (THA) ceramic-on-ceramic (CoC) bearings are a potential for overcoming the wear that is seen in ceramic-on-polyethylene (CoPE) bearings, and can lead to wear-induced osteolysis, resulting in loosening of the implant. However, CoC bearings show disadvantages as well, such as squeaking sounds and being more fragile, which can cause ceramic head or liner fracture. Because comparative long-term studies are limited, the objective of this study was to determine the long-term difference in wear, identify potential predictive factors for wear, investigate radiological findings such as osteolysis, and evaluate clinical functioning and complications between these bearings. AIM: To determine 10-year differences in wear, predictive factors for wear, and investigate radiological findings and clinical functioning between CoC and CoPE. METHODS: This observational prospective single-center cohort study with a 10-year follow-up includes a documented series of elective THAs. Primary outcome was wear measured by anteroposterior (AP) radiographs. Secondary outcomes were potential predictive factors for wear, complications during follow-up, Harris hip score (HHS), and radiological findings such as presence of radiolucency, osteolysis, atrophy, and hypertrophy around the cup. Due to the absence of wear in the CoC group, stratified analysis to identify risk factors for wear was only performed in the CoPE group by use of univariate linear regression analysis. HHS was expressed as a change from baseline and the association with bearing type was assessed by use of multivariate linear regression analysis, adjusted for potential confounders. RESULTS: = 0.99). No significant differences in complications and radiological findings were seen between groups. CONCLUSION: CoC bearing shows lower wear rates compared to CoPE at 10-year follow-up with cup inclination as a predictive factor for wear and no differences in complications, HHS, and radiological findings.

Topics & Concepts

MedicineCeramicComposite materialDentistryMaterials scienceOrthopaedic implants and arthroplastyTotal Knee Arthroplasty OutcomesOrthopedic Infections and Treatments