Litcius/Paper detail

A Novel Device Model Validation Strategy for 1.5- and 3-T MRI Heating Safety Assessment

Yu Wang, Shuo Song, Jianfeng Zheng, Qingyan Wang, Stuart A. Long, Wolfgang Kainz, Ji Chen

2020IEEE Transactions on Instrumentation and Measurement23 citationsDOI

Abstract

This article presents a novel test instrument and strategy for active implantable medical device (AIMD) model validation under a radiofrequency (RF) field generated by a magnetic resonance imaging (MRI) system. A high electric field generator is developed as an alternative instrument for standard MRI RF coils. This system is both compact and power efficient. In addition, a set of validation pathways are proposed based on the Hadamard matrix. These pathways provide low correlations among incident electric fields and yet generate relatively high temperature rises. Two generic AIMDs with different lengths are used to validate the strategy experimentally. It is demonstrated that this method provides an efficient and effective alternative for future AIMD model validation under MRI RF emissions.

Topics & Concepts

Radio frequencyGenerator (circuit theory)Computer scienceElectronic engineeringField (mathematics)Hadamard transformMagnetic resonance imagingSet (abstract data type)Reliability engineeringPower (physics)EngineeringPhysicsTelecommunicationsMathematicsMedicineRadiologyPure mathematicsProgramming languageQuantum mechanicsUltrasonics and Acoustic Wave PropagationMicrowave and Dielectric Measurement TechniquesWireless Power Transfer Systems