Dosimetry of ultra-high dose rate electron beams using thermoluminescence and optically stimulated luminescence detectors
Silvia Motta, Riccardo Dal Bello, Jeppe Brage Christensen, Lily Bossin, E.G. Yukihara
Abstract
Abstract Objective. The aim of this work is to investigate the dose rate dependence of thermoluminescence and optically stimulated luminescence detectors (TLDs and OSLDs) in a wide uniform ultra-high dose rate electron beam and demonstrate the potential use of TLDs and OSLDs to correct the ion recombination in air-filled ionization chambers. This study avoids previously reported complications related to the field size and homogeneity. Approach. Two types of OSLDs (BeO and Al 2 O 3 :C) and three types of TLDs (LiF:Mg,Ti, LiF:Mg,Cu,P, CaF 2 :Tm) were irradiated simultaneously in a uniform 16 MeV electron beam generated by a clinically decommissioned C-Arm LINAC, modified to deliver doses per pulse between 8.3 × 10 −4 Gy and 1.255 Gy, corresponding to instantaneous dose rates between 2 × 10 2 Gy s −1 and 3 × 10 5 Gy s −1 . A prototype ultra-thin parallel plate ionization chamber was employed as reference detector. Main results. Reproducible results were achieved both at conventional (standard deviation of the data <2%) and at the highest dose per pulse (standard deviation of the data <4%). No trend in the dose rate response of the TLDs and OSLDs was observed in the investigated dose per pulse range. The Al 2 O 3 :C OSLD was found to be the most precise detector, with a standard deviation of the data <2% at all investigated dose rates and dose levels. Significance. The dose rate independence of the investigated TLDs and OSLDs make them good candidates for dosimetry at ultra-high dose rates, at least up to 3 × 10 5 Gy s −1 . A dose rate independent method to measure the dose per pulse is proposed, which can be applied to characterize ultra-high dose rate electron beams and correct for ion recombination in ionization chambers.