The accuracy of a novel image‐guided hybrid robotic system for dental implant placement: An in vitro study

Baoxin Tao, Yuan Feng, Xingqi Fan, Kengliang Lan, Minjie Zhuang, Shigang Wang, Feng Wang, Xiaojun Chen, Yiqun Wu

2022International Journal of Medical Robotics and Computer Assisted Surgery44 citationsDOI

Abstract

BACKGROUND: This in vitro study aims to evaluate the accuracy of dental implant placement by a novel image-guided hybrid robotic system for dental implant surgery (HRS-DIS). METHODS: The HRS-DIS with a 5 degree of freedom (DOF) serial manipulator and a 6 DOF Stewart platform was developed. To evaluate the accuracy of repeated drilling, the holes were prepared twice with a 2.2 mm drill. To evaluate the accuracy of dental implant placement, the entry, exit and angle deviations of dental implants were measured. RESULTS: Twenty-four holes were prepared twice, and mean (±SD) of diameters were measured as 2.2 ± 0.02 mm. A total of 160 dental implants were placed in 32 phantoms by HRS-DIS. The mean (±SD) of the entry, exit and angle deviation were 0.8 ± 0.54 mm, 0.87 ± 0.54 mm and 1.0 1 ± 0.44°, respectively. CONCLUSIONS: The results of the in vitro study preliminarily validated that the HRS-DIS could provide a high accuracy for dental implant surgery.