FT-ICR Mass Spectrometry Imaging at Extreme Mass Resolving Power Using a Dynamically Harmonized ICR Cell with 1ω or 2ω Detection

Mathieu Tiquet, Raphaël La Rocca, Stefan Kirnbauer, Samuele Zoratto, Daan van Kruining, Loïc Quinton, Gauthier Eppe, Pilar Martínez‐Martínez, Martina Marchetti‐Deschmann, Edwin De Pauw, Johann Far

2022Analytical Chemistry28 citationsDOIOpen Access PDF

Abstract

shift. Adjustment of the laser parameters is proposed to maintain high spectral resolution and mass accuracy measurement within the instrument specifications during MSI analyses. The optimized method has been successfully employed in replicates to perform high-quality MALDI MS images at resolving power (FWHM) above 1,000,000 in the lipid mass range across the whole image for superconducting magnets of 7T and 9.4T using 1 and 2ω detection. Our data also compare favorably with MALDI MSI experiments performed on higher-magnetic-field superconducting magnets, including the 21T MALDI FT-ICR prototype instrument of the NHMFL group at Tallahassee, Florida.