Agreement between swept-source optical biometry and Scheimpflug-based topography measurements of posterior corneal curvature

Bülent Köse

2021Journal of Cataract & Refractive Surgery21 citationsDOI

Abstract

PURPOSE: To evaluate agreement of posterior corneal curvature parameters between a swept-source optical coherence tomography-based biometer (IOLMaster 700) and Scheimpflug topography (Pentacam HR) in healthy, myopic eyes. SETTING: Department of Ophthalmology, Osmangazi Aritmi Hospital, Bursa, Turkey. DESIGN: Retrospective study. METHODS: In this study, 59 eyes of 59 patients who applied as candidates for refractive surgery were included. Measurements of the posterior corneal curvature were performed using IOLMaster 700 and Pentacam HR. J0 and J45 vector components were calculated using astigmatic values. Posterior corneal keratometry (K) measurements at flat (K flat) and steep (K steep) axes and J0 and J45 values were compared between the 2 devices. The agreement between the devices was analyzed using the Bland-Altman method. Intraclass correlation coefficients and within-subject SD were calculated to assess the repeatability. RESULTS: 59 eyes of 59 patients were included in this study. The IOLMaster 700 provided significantly flatter K flat and K steep values (P < .001, for both). Significant differences were found in J0 and J45 values (P = 0.13 and P < .001, respectively). The mean differences between K flat, K steep, and J0 and J45 values were 0.49 diopter (D), 0.53 D, 0.04 D, and -0.05 D, respectively. CONCLUSIONS: The IOLMaster 700 and the Pentacam HR could not be used interchangeably to measure K flat, K steep, and J0 and J45 values of the posterior corneal curvature in healthy, myopic eyes. Both devices showed high repeatability for posterior corneal curvature measurement.

Topics & Concepts

Scheimpflug principleKeratometerOphthalmologyRepeatabilityDioptreCurvatureMedicinePhysicsMathematicsOpticsCorneaGeometryVisual acuityStatisticsOphthalmology and Visual Impairment StudiesCorneal surgery and disordersGlaucoma and retinal disorders
Agreement between swept-source optical biometry and Scheimpflug-based topography measurements of posterior corneal curvature | Litcius