A Comparison of Wearable Tonometry, Photoplethysmography, and Electrocardiography for Cuffless Measurement of Blood Pressure in an Ambulatory Setting

Rebecca Mieloszyk, Hope Twede, Jonathan Lester, Jeremiah Wander, Sumit Basu, Gabe Cohn, Greg Smith, Dan Morris, Sidhant Gupta, Desney Tan, Nicolas Villar, Moni Wolf, Sailaja Malladi, Matt Mickelson, Lauren Ryan, Lindsey Kim, Jeffrey Kepple, Susanne Kirchner, Emma Wampler, Riena Terada, Joel Robinson, Ron Paulsen, T. Scott Saponas

2022IEEE Journal of Biomedical and Health Informatics74 citationsDOI

Abstract

OBJECTIVE: While non-invasive, cuffless blood pressure (BP) measurement has demonstrated relevancy in controlled environments, ambulatory measurement is important for hypertension diagnosis and control. We present both in-lab and ambulatory BP estimation results from a diverse cohort of participants. METHODS: Participants (N=1125, aged 21-85, 49.2% female, multiple hypertensive categories) had BP measured in-lab over a 24-hour period with a subset also receiving ambulatory measurements. Radial tonometry, photoplethysmography (PPG), electrocardiography (ECG), and accelerometry signals were collected simultaneously with auscultatory or oscillometric references for systolic (SBP) and diastolic blood pressure (DBP). Predictive models to estimate BP using a variety of sensor-based feature groups were evaluated against challenging baselines. RESULTS: Despite limited availability, tonometry-derived features showed superior performance compared to other feature groups and baselines, yieldingprediction errors of 0.32 ±9.8 mmHg SBP and 0.54 ±7.7 mmHg DBP in-lab, and 0.86 ±8.7 mmHg SBP and 0.75 ±5.9 mmHg DBP for 24-hour averages. SBP error standard deviation (SD) was reduced in normotensive (in-lab: 8.1 mmHg, 24-hr: 7.2 mmHg) and younger (in-lab: 7.8 mmHg, 24-hr: 6.7 mmHg) subpopulations. SBP SD was further reduced 15-20% when constrained to the calibration posture alone. CONCLUSION: Performance for normotensive and younger participants was superior to the general population across all feature groups. Reference type, posture relative to calibration, and controlled vs. ambulatory setting all impacted BP errors. SIGNIFICANCE: Results highlight the need for demographically diverse populations and challenging evaluation settings for BP estimation studies. We present the first public dataset of ambulatory tonometry and cuffless BP over a 24-hour period to aid in future cardiovascular research.

Topics & Concepts

MedicineAmbulatoryWearable computerBlood pressureElectrocardiographyAmbulatory blood pressurePhotoplethysmogramCardiologyInternal medicineRemote patient monitoringWearable technologyEmergency medicineTelemedicineEstimationMedical emergencyPhysical therapyIntensive care medicineNon-Invasive Vital Sign MonitoringHeart Rate Variability and Autonomic ControlOptical Imaging and Spectroscopy Techniques