Accuracy and Systematic Biases of Heart Rate Measurements by Consumer-Grade Fitness Trackers in Postoperative Patients: Prospective Clinical Trial
Philipp Helmer, Sebastian Hottenrott, Philipp Rodemers, Robert Leppich, Maja Helwich, Rüdiger Pryss, Peter Kranke, Patrick Meybohm, Bernd Winkler, Michael Sammeth
Abstract
BACKGROUND: Over the recent years, technological advances of wrist-worn fitness trackers heralded a new era in the continuous monitoring of vital signs. So far, these devices have primarily been used for sports. OBJECTIVE: However, for using these technologies in health care, further validations of the measurement accuracy in hospitalized patients are essential but lacking to date. METHODS: We conducted a prospective validation study with 201 patients after moderate to major surgery in a controlled setting to benchmark the accuracy of heart rate measurements in 4 consumer-grade fitness trackers (Apple Watch 7, Garmin Fenix 6 Pro, Withings ScanWatch, and Fitbit Sense) against the clinical gold standard (electrocardiography). RESULTS: ≥0.94) coefficients, with a relative error as low as mean absolute percentage error <5% based on 1630 valid measurements. We identified confounders significantly biasing the measurement accuracy, although not at clinically relevant levels (mean absolute error<5 beats per minute). CONCLUSIONS: Consumer-grade fitness trackers appear promising in hospitalized patients for monitoring heart rate. TRIAL REGISTRATION: ClinicalTrials.gov NCT05418881; https://www.clinicaltrials.gov/ct2/show/NCT05418881.