Litcius/Paper detail

Use of Electrical Impedance Tomography (EIT) to Estimate Tidal Volume in Anaesthetized Horses Undergoing Elective Surgery

Benedetta Crivellari, Anthea L. Raisis, Giselle Hosgood, Andreas D. Waldmann, David J. Murphy, Martina Mosing

2021Animals18 citationsDOIOpen Access PDF

Abstract

This study explores the application of electric impedance tomography (EIT) to estimate tidal volume (VT) by measuring impedance change per breath (∆Zbreath). Seventeen healthy horses were anaesthetised and mechanically ventilated for elective procedures requiring dorsal recumbency. Spirometric VT (VTSPIRO) and ∆Zbreath were recorded periodically; up to six times throughout anaesthesia. Part 1 assessed these variables at incremental delivered VT of 10, 12 and 15 mL/kg. Part 2 estimated VT (VTEIT) in litres from ∆Zbreath at three additional measurement points using a line of best fit obtained from Part 1. During part 2, VT was adjusted to maintain end-tidal carbon dioxide between 45–55 mmHg. Linear regression determined the correlation between VTSPIRO and ∆Zbreath (part 1). Estimated VTEIT was assessed for agreement with measured VTSPIRO using Bland Altman analysis (part 2). Marked variability in slope and intercepts was observed across horses. Strong positive correlation between ∆Zbreath and VTSPIRO was found in each horse (R2 0.9–0.99). The agreement between VTEIT and VTSPIRO was good with bias (LOA) of 0.26 (−0.36–0.88) L. These results suggest that, in anaesthetised horses, EIT can be used to monitor and estimate VT after establishing the individual relationship between these variables.

Topics & Concepts

Electrical impedance tomographyLimits of agreementTidal volumeMedicineNuclear medicineHorseAnesthesiaImpedance cardiographyVolume (thermodynamics)TomographyAnatomyRespiratory systemInternal medicineRadiologyPhysicsHeart rateGeologyStroke volumeQuantum mechanicsPaleontologyBlood pressureVeterinary Equine Medical ResearchFlow Measurement and AnalysisHemodynamic Monitoring and Therapy
Use of Electrical Impedance Tomography (EIT) to Estimate Tidal Volume in Anaesthetized Horses Undergoing Elective Surgery | Litcius