Litcius/Paper detail

Short single-wedge stems have higher risk of periprosthetic fracture than other cementless stem designs in Dorr type A femurs: a finite element analysis

Sumon Nandi, Anoli Shah, Amin Joukar, Kevin Becker, Cody Crutchfield, Vijay K. Goel

2020Hip International14 citationsDOI

Abstract

BACKGROUND: The use of total hip arthroplasty (THA) femoral stems that transmit force in a favourable manner to the femur may minimise periprosthetic fractures. Finite element analysis (FEA) is a computerised method that analyses the effect of forces applied to a structure with complex shape. Our aim was to apply FEA to compare primary THA cementless stem designs and their association with periprosthetic fracture risk. METHODS: 3-dimensional (3D) models of a Dorr Type A femur and 5 commonly used primary THA cementless stem designs (short single wedge, standard-length single wedge, modular, double-wedge metaphyseal filling, and cylindrical fully coated) were developed using computed tomography (CT) imaging. Implant insertion, single-leg stance, and twisting with a planted foot were simulated. FEA was performed, and maximum femoral strain along the implant-bone interface recorded. RESULTS: Femoral strain was highest with short single-wedge stem design (0.3850) and lowest with standard-length single-wedge design (0.0520). Location of maximum femoral strain varied by stem design, but not with implant insertion, single-leg stance, or twisting with a planted foot. Strain was as high during implant insertion as with single-leg stance or twisting with a planted foot. CONCLUSIONS: Our results suggest the risk of intraoperative and postoperative periprosthetic fracture with THA in a Dorr A femur is highest with short single-wedge stems and lowest with standard-length single-wedge stems. Consideration may be given to minimising the use of short single-wedge stems in THA. Implant-specific sites of highest strain should be carefully inspected for fracture.

Topics & Concepts

PeriprostheticWedge (geometry)ImplantFemurMedicineFinite element methodFemur fractureOrthodonticsFemoral fractureArthroplastySurgeryStructural engineeringGeometryMathematicsEngineeringOrthopaedic implants and arthroplastyTotal Knee Arthroplasty OutcomesHip and Femur Fractures
Short single-wedge stems have higher risk of periprosthetic fracture than other cementless stem designs in Dorr type A femurs: a finite element analysis | Litcius