A Highly Versatile X-ray and Electron Beam Diamond Dosimeter for Radiation Therapy and Protection
Sara Pettinato, M. Girolami, A. Stravato, Valerio Serpente, Daniela Musio, M.C. Rossi, D.M. Trucchi, Riccardo Olivieri, S. Salvatori
Abstract
Radiotherapy is now recognized as a pillar in the fight against cancer. Two different types are currently used in clinical practice: (1) external beam radiotherapy, using high-energy X-rays or electron beams, both in the MeV-range, and (2) intraoperative radiotherapy, using low-energy X-rays (up to 50 keV) and MeV-range electron beams. Versatile detectors able to measure the radiation dose independently from the radiation nature and energy are therefore extremely appealing to medical physicists. In this work, a dosimeter based on a high-quality single-crystal synthetic diamond sample was designed, fabricated and characterized under low-energy X-rays, as well as under high-energy pulsed X-rays and electron beams, demonstrating excellent linearity with radiation dose and dose-rate. Detector sensitivity was measured to be 0.299 ± 0.002 µC/Gy under 6 MeV X-ray photons, and 0.298 ± 0.004 µC/Gy under 6 MeV electrons, highlighting that the response of the diamond dosimeter is independent of the radiation nature. Moreover, in the case of low-energy X-rays, an extremely low limit of detection (23 nGy/s) was evaluated, pointing out the suitability of the device to radiation protection dosimetry.