Fabrication of Silver Nanowire/Polydimethylsiloxane Dry Electrodes by a Vacuum Filtration Method for Electrophysiological Signal Monitoring

Ravinder Reddy Kisannagar, Pallavi Jha, Ambuja Navalkar, Samir K. Maji, Dipti Gupta

2020ACS Omega69 citationsDOIOpen Access PDF

Abstract

Flexible and dry electrodes have attracted huge attention due to their potential application in long-term electrophysiological signal monitoring. In this work, we present a novel method to pattern silver nanowires (AgNWs) on a polydimethylsiloxane (PDMS) substrate-based dry electrodes by a vacuum filtration method for electrophysiological signal monitoring. The Scotch tape peel-off test confirms the excellent adhesion of the patterned AgNWs on a PDMS substrate. The cytotoxicity of the proposed electrode is detected by an MTT assay method, and 90% cell viability is observed for the period of one week, indicating no cytotoxic effect on living cells. The signal to noise ratios of the conventional wet Ag/AgCl and dry AgNW/PDMS electrodes are 24.6 and 25.4 dB, indicating that AgNW/PDMS dry electrodes measure a high-quality electrophysiological signal when compared with that of the conventional Ag/AgCl wet electrodes.

Topics & Concepts

PolydimethylsiloxaneElectrodeMaterials scienceSubstrate (aquarium)SIGNAL (programming language)Filtration (mathematics)ElectrophysiologyNanotechnologyAdhesionOptoelectronicsFabricationComposite materialChemistryPathologyInternal medicineProgramming languageGeologyComputer scienceMathematicsAlternative medicinePhysical chemistryMedicineStatisticsOceanographyAdvanced Sensor and Energy Harvesting MaterialsNeuroscience and Neural EngineeringConducting polymers and applications