Litcius/Paper detail

A Sensitive and Specific Liquid Chromatography-Tandem Mass Spectrometry Assay for Simultaneous Quantification of Salivary Melatonin and Cortisol: Development and Comparison With Immunoassays

Sunghwan Shin, 오현주, Hea Ree Park, Eun Yeon Joo, Soo‐Youn Lee

2020Annals of Laboratory Medicine26 citationsDOIOpen Access PDF

Abstract

Melatonin and cortisol are clinically important for diagnosing sleep and mood disorders. We developed and validated a liquid chromatography-tandem mass spectrometry (LC-MS/MS) assay for simultaneous measurement of salivary melatonin and cortisol concentrations according to the Clinical and Laboratory Standards Institute guidelines. Additionally, we compared the LC-MS/MS assay with immunoassays, ELISA (Direct Salivary Melatonin Elisa EK-DSM, Bühlmann Laboratories AG, Schönenbuch, Switzerland) and electrochemiluminescence immunoassay (Cortisol II, Roche, Mannheim, Germany), using 121 saliva samples. The LC-MS/MS assay exhibited good performance in terms of linearity, precision, accuracy, limit of detection, lower limit of quantification, extraction recovery, carry-over, and matrix effect. The LC-MS/MS assay and immunoassays showed strong correlation (Pearson's r=0.910 for melatonin, r=0.955 for cortisol), but demonstrated a significant mean bias of 23.2% (range 54.0-143.7%) for melatonin and 48.9% (range 59.7-184.7%) for cortisol. Our LC-MS/MS assay provided more sensitive and reliable salivary melatonin and cortisol quantification results compared with immunoassays.

Topics & Concepts

MelatoninChromatographyImmunoassayChemistryDetection limitLiquid chromatography–mass spectrometrySalivaTandem mass spectrometryMass spectrometryInternal medicineMedicineImmunologyBiochemistryAntibodyStress Responses and CortisolTryptophan and brain disordersCircadian rhythm and melatonin