Variability in haemoglobin concentration by measurement tool and blood source: an analysis from seven countries
Aviva I Rappaport, Crystal D Karakochuk, Sonja Y. Hess, Ralph D. Whitehead, Sorrel Namasté, Omar Dary, Megan Parker, Lynnette M. Neufeld, Leila M Larson, Sam Newton, Rita Wegmüller, Denish Moorthy
Abstract
OBJECTIVE: We explore factors such as the blood sampling site (capillary vs venous), the equipment (HemoCue vs automated haematology analyser) and the model of the HemoCue device (201+ vs 301) that may impact haemoglobin measurements in capillary and venous blood. METHODS: Eleven studies were identified, and bias, concordance and measures of diagnostic performance were assessed within each study. FINDINGS: Our analysis included 11 studies from seven countries (Cambodia, India, The Gambia, Ghana, Laos, Rwanda and USA). Samples came from children, men, non-pregnant women and pregnant women. Mean bias ranged from -8.7 to 2.5 g/L in Cambodian women, 6.2 g/L in Laotian children, 2.4 g/L in Ghanaian women, 0.8 g/L in Gambian children 6-23 months and 1.4 g/L in Rwandan children 6-59 months when comparing capillary blood on a HemoCue to venous blood on a haematology analyser. Bias was 8.3 g/L in Indian non-pregnant women and 2.6 g/L in Laotian children and women and 1.5 g/L in the US population when comparing capillary to venous blood using a HemoCue. For venous blood measured on the HemoCue compared with the automated haematology analyser, bias was 5.3 g/L in Gambian pregnant women 18-45 years and 11.3 g/L in Laotian children 6-59 months. CONCLUSION: Our analysis found large variability in haemoglobin concentration measured on capillary or venous blood and using HemoCue Hb 201+ or Hb 301 or automated haematology analyser. We cannot ascertain whether the variation is due to differences in the equipment, differences in capillary and venous blood, or factors affecting blood collection techniques.