A Single-Center Experience with the Olympus ORBEYE 4K-3D Exoscope for Microsurgery of Complex Cranial Cases: Technical Nuances and Learning Curve
Giovanni Muscas, Francesca Battista, Andrea Boschi, Francesca Morone, Alessandro Della Puppa
Abstract
Abstract Background The introduction of exoscopes in neurosurgery has been welcomed due to their maneuverability, ergonomics, and low-profile frame. 3D devices have further enabled a better stereoscopic visualization. Reports on their application, albeit more and more frequent, are still at their beginning stages. We present our experience with the Olympus ORBEYE 4K-3D exoscope for major cranial procedures. The strengths and weaknesses of the exoscope are presented, and the nuances associated with the learning curve are illustrated. Methods Over 2 weeks, patients undergoing surgery for major cranial pathologies were offered to participate in this evaluation of the Olympus ORBEYE 4K-3D exoscope. Information on the use of the exoscope was collected to assess the features and struggles in the learning curve. A comparison with the operating microscope was made. Results Fourteen patients with different intracranial pathologies were operated on with the exoscope. No surgery-related complications occurred. The microsurgical part was performed with the exoscope in six cases. The exoscope was used for 72.9% (±37.5%) of the whole microsurgical time vs. 27.1% (±37.5%) microscope time (p = 0.02). Conclusion The Olympus ORBEYE 4K-3D exoscope represents a useful evolution of the operating microscope. It requires time to overcome potential difficulties, mostly related to previous motor schemes acquired with operating microscopes. Its features could represent the basis for a paradigm shift in microsurgery.