Litcius/Paper detail

Integrating electrical impedance tomography and transpulmonary pressure monitoring to personalize PEEP in hypoxemic patients undergoing pressure support ventilation

Douglas Slobod, Marco Leali, Elena Spinelli, Domenico Luca Grieco, Savino Spadaro, Tommaso Mauri

2022Critical Care22 citationsDOIOpen Access PDF

Abstract

Monitoring with electrical impedance tomography (EIT) during a decremental PEEP trial has been used to identify the PEEP that yields the optimal balance of pulmonary overdistension and collapse. This method is based on pixel-level changes in respiratory system compliance and depends on fixed or measured airway driving pressure. We developed a novel approach to quantify overdistension and collapse during pressure support ventilation (PSV) by integrating transpulmonary pressure and EIT monitoring and performed pilot tests in three hypoxemic patients. We report that our experimental approach is feasible and capable of identifying a PEEP that balances overdistension and collapse in intubated hypoxemic patients undergoing PSV.

Topics & Concepts

MedicineElectrical impedance tomographyMechanical ventilationVentilation (architecture)Intensive care medicineCardiologyEmergency medicineAnesthesiaTomographyRadiologyMechanical engineeringEngineeringRespiratory Support and MechanismsNon-Invasive Vital Sign MonitoringHemodynamic Monitoring and Therapy