Litcius/Paper detail

Pulse-sampling fluorescence lifetime imaging: evaluation of photon economy

Xiangnan Zhou, Julien Bec, Katjana Ehrlich, Alba Alfonso‐García, Laura Marcu

2023Optics Letters15 citationsDOIOpen Access PDF

Abstract

This Letter presents an experimental study comparing the photon rate and photon economy of pulse sampling fluorescence lifetime imaging (PS-FLIm) with the conventional time-correlated single photon counting (TCSPC) technique. We found that PS-FLIm has a significantly higher photon detection rate (200 MHz) compared with TCSPC (2-8 MHz) but lower photon economy (4-5 versus 1-1.3). The main factor contributing to the lower photon economy in PS-FLIm is laser pulse variability. These results demonstrate that PS-FLIm offers 25× faster imaging speed than TCSPC while maintaining room light rejection in clinical settings. This makes PS-FLIm a robust technique for clinical applications.

Topics & Concepts

Fluorescence-lifetime imaging microscopyPhotonPulse (music)Photon countingFluorescenceMaterials scienceOpticsSampling (signal processing)Two-photon excitation microscopyLaserPhysicsDetectorOptical Imaging and Spectroscopy TechniquesAdvanced Optical Sensing TechnologiesAdvanced Fluorescence Microscopy Techniques