Inflated Gamma Interferon Response with QuantiFERON-TB Gold Plus Using the Automated Liaison XL Analyzer: a Testing Algorithm To Mitigate False-Positive Results in Low-Incidence Settings
Vladimir Buron, Niaz Banaei
Abstract
< 0.0001). The CLIA had positive and negative percent agreement levels with the ELISA of 91.7% (121/132) and 95.2% (139/146), respectively. In borderline-negative samples tested with ELISA, CLIA was positive in 42.7% (94/220). CLIA with a standard curve resulted in 36.4% (80/220) positivity. Retesting CLIA false positives (TB1 or TB2 range, 0 to ≤1.3 IU/mL) with ELISA reduced false positives by 84.3% (59/70). Retesting with CLIA reduced the false-positive rate by 10.4% (8/77). Implementing the Liaison CLIA for QFT-Plus in low-incidence settings risks falsely elevating conversion rates and overburdening clinics and overtreating patients. Confirming borderline positives with ELISA is a viable strategy to mitigate false-positive CLIA results.