BACKGROUNDMitral regurgitation (MR) determines pathophysiology and outcome in advanced heart failure. The impact of left ventricular assist device (LVAD) placement on clinically significant MR and its contribution to long-term outcomes has been sparsely evaluated.METHODSWe evaluated the effect of clinically significant MR on patients implanted in the MOMENTUM 3 trial with either the HeartMate II (HMII) or the HeartMate 3 (HM3) at 2 years. Clinical significance was defined as moderate or severe grade MR determined by site-based echocardiograms.RESULTSOf 927 patients with LVAD implants without a prior or concomitant mitral valve procedure, 403 (43.5%) had clinically significant MR at baseline. At 1-month of support, residual MR was present in 6.2% of patients with HM3 and 14.3% of patients with HMII (relative risk = 0.43; 95% CI, 0.22–0.84; p = 0.01) with a low rate of worsening at 2 years. Residual MR at 1-month post-implant did not impact 2-year mortality for either the HM3 (hazard ratio [HR],1.41; 95% CI, 0.52–3.89; p = 0.50) or HMII (HR, 0.91; 95% CI, 0.37–2.26; p = 0.84) LVAD. The presence or absence of baseline MR did not influence mortality (HM3 HR, 0.86; 95% CI, 0.56–1.33; p = 0.50; HMII HR, 0.81; 95% CI, 0.54–1.22; p = 0.32), major adverse events or functional capacity. In multivariate analysis, severe baseline MR (p = 0.001), larger left ventricular dimension (p = 0.002), and implantation with the HMII instead of the HM3 LVAD (p = 0.05) were independently associated with an increased likelihood of persistent MR post-implant.CONCLUSIONSHemodynamic unloading after LVAD implantation improves clinically significant MR early, sustainably, and to a greater extent with the HM3 LVAD. Neither baseline nor residual MR influence outcomes after LVAD implantation. Mitral regurgitation (MR) determines pathophysiology and outcome in advanced heart failure. The impact of left ventricular assist device (LVAD) placement on clinically significant MR and its contribution to long-term outcomes has been sparsely evaluated. We evaluated the effect of clinically significant MR on patients implanted in the MOMENTUM 3 trial with either the HeartMate II (HMII) or the HeartMate 3 (HM3) at 2 years. Clinical significance was defined as moderate or severe grade MR determined by site-based echocardiograms. Of 927 patients with LVAD implants without a prior or concomitant mitral valve procedure, 403 (43.5%) had clinically significant MR at baseline. At 1-month of support, residual MR was present in 6.2% of patients with HM3 and 14.3% of patients with HMII (relative risk = 0.43; 95% CI, 0.22–0.84; p = 0.01) with a low rate of worsening at 2 years. Residual MR at 1-month post-implant did not impact 2-year mortality for either the HM3 (hazard ratio [HR],1.41; 95% CI, 0.52–3.89; p = 0.50) or HMII (HR, 0.91; 95% CI, 0.37–2.26; p = 0.84) LVAD. The presence or absence of baseline MR did not influence mortality (HM3 HR, 0.86; 95% CI, 0.56–1.33; p = 0.50; HMII HR, 0.81; 95% CI, 0.54–1.22; p = 0.32), major adverse events or functional capacity. In multivariate analysis, severe baseline MR (p = 0.001), larger left ventricular dimension (p = 0.002), and implantation with the HMII instead of the HM3 LVAD (p = 0.05) were independently associated with an increased likelihood of persistent MR post-implant. Hemodynamic unloading after LVAD implantation improves clinically significant MR early, sustainably, and to a greater extent with the HM3 LVAD. Neither baseline nor residual MR influence outcomes after LVAD implantation.

Topics & Concepts

MedicineCardiologyInternal medicineMitral regurgitationHazard ratioImplantConcomitantHeart failureFunctional mitral regurgitationVentricular assist deviceDestination therapySurgeryEjection fractionConfidence intervalMechanical Circulatory Support DevicesCardiac Structural Anomalies and RepairCardiac and Coronary Surgery Techniques
Impact of left ventricular assist device implantation on mitral regurgitation: An analysis from the MOMENTUM 3 trial | Litcius