Martijn J. H. Doeleman, Anouk Esseveld, Albert Huisman, Sytze de Roock, Wouter M. Tiel Groenestege

2023International Journal of Laboratory Hematology26 citationsDOIOpen Access PDF

Abstract

INTRODUCTION: This study assessed the comparability of complete blood count (CBC) parameters between capillary and venous samples, and extended previous research by examining the influence of different storage temperatures on CBC stability up to 7 days after sample collection. METHODS: Venous and capillary blood samples were collected from 93 adult patients. Hemoglobin (Hb), hematocrit (Ht), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), mean corpuscular hemoglobin concentration (MCHC), mean platelet volume (MPV), leukocytes, lymphocytes, basophils, eosinophils, erythrocytes, red cell distribution width (RDW), immature granulocytes (IG), immature reticulocyte fraction (IRF), monocytes, neutrophils, platelets, and reticulocytes were measured. Deming regression and mean relative differences between venous and capillary measurements were contrasted with desirable total allowable error (TEa). Stability was assessed in 20-27 venous blood samples stored at 4, 21-22, or 30°C, and analyzed at 0, 24, 48, 72, 96, 120, 144, and 168 h. Mean relative change with respect to baseline measurements was compared to the desirable TEa to determine acceptable stability. RESULTS: Deming regression demonstrated strong linear correlations and acceptable variation between venous and capillary measurements. Erythrocytes, Hb, Ht, MCH, MCV, RDW, reticulocytes, and platelets showed acceptable stability for at least 96 h at 4°C. Mean relative change exceeded desirable TEa after 24 h at 30°C for all parameters, except erythrocytes, Hb, leukocytes, and MCH. CONCLUSION: Clinical laboratory specialists and clinicians should be aware of potential differences between venous and capillary measurements, and the influence of storage conditions. Clinical validity of delayed CBC analysis depends on the clinical situation and required precision of the result.

Topics & Concepts

Mean corpuscular volumeHematocritMean corpuscular hemoglobinHemoglobinVenous bloodPlateletMean corpuscular hemoglobin concentrationMedicineLinear regressionRed blood cell distribution widthReticulocyteBlood volumeInternal medicineChemistryGastroenterologyBiochemistryMathematicsGeneMessenger RNAStatisticsClinical Laboratory Practices and Quality ControlInflammatory Biomarkers in Disease PrognosisBiosimilars and Bioanalytical Methods
Stability and comparison of complete blood count parameters between capillary and venous blood samples | Litcius