Ultra-low foetal radiation exposure in 18F-FDG PET/CT imaging with a long axial field-of-view PET/CT system

Charlotte L. C. Smith, Maqsood Yaqub, R.H.H. Wellenberg, Jelijn J. Knip, Ronald Boellaard, Gerben J.C. Zwezerijnen

2024EJNMMI Physics15 citationsDOIOpen Access PDF

Abstract

Abstract Purpose Long axial field-of-view (LAFOV) PET/CT systems enable PET/CT scans with reduced injected activities because of improved sensitivity. With this study, we aimed to examine the foetal radiation dose from an 18 F-FDG PET/CT scan on a LAFOV PET/CT system with reduced injected activity. Methods Two pregnant women were retrospectively included and received an 18 F-FDG PET/CT scan on a LAFOV PET/CT system with an intravenous bolus injection of 0.30 MBq/kg. Foetal radiation exposure from the PET was estimated using dose conversion factors from three published papers. Radiation exposure from the CT scans was estimated using CT-Expo. Results Foetal radiation dose from the PET scan ranged between 0.11 and 0.44 mGy. Foetal radiation exposure from the CT scan ranged between < 0.10 – 0.90 mGy depending if the foetus was included in the field-of-view. Conclusion Foetal radiation dose could be reduced to < 1.5 mGy when scanning pregnant patients on a LAFOV PET/CT system. The radiation dose to the foetus was reduced significantly in our study due to the increased sensitivity of the LAFOV PET/CT system.