Volumetric computed tomography with carbon nanotube X-ray source array for improved image quality and accuracy

Shuang Xu, Yuanming Hu, Boyuan Li, Christina R. Inscoe, Donald A. Tyndall, Yueh Z. Lee, Jianping Lü, Otto Zhou

2023Communications Engineering13 citationsDOIOpen Access PDF

Abstract

Cone beam computed tomography (CBCT) is widely used in medical and dental imaging. Compared to a multidetector CT, it provides volumetric images with high isotropic resolution at a reduced radiation dose, cost and footprint without the need for patient translation. The current CBCT has several intrinsic limitations including reduced soft tissue contrast, inaccurate quantification of X-ray attenuation, image distortions and artefacts, which have limited its clinical applications primarily to imaging hard tissues and made quantitative analysis challenging. Here we report a multisource CBCT (ms-CBCT) which overcomes the short-comings of the conventional CBCT by using multiple narrowly collimated and rapidly scanning X-ray beams from a carbon nanotube field emission source array. Phantom imaging studies show that, the ms-CBCT increases the accuracy of the Hounsfield unit values by 60%, eliminates the cone beam artefacts, extends the axial coverage, and improves the soft tissue contrast-to-noise ratio by 30-50%, compared to the CBCT configuration.

Topics & Concepts

Cone beam computed tomographyHounsfield scaleImaging phantomCollimated lightImage qualityScannerTomosynthesisImage-guided radiation therapyAttenuationMedical imagingMaterials scienceTomographyNuclear medicineMedical physicsComputed tomographyOpticsMedicineRadiologyComputer sciencePhysicsComputer visionImage (mathematics)MammographyBreast cancerLaserCancerInternal medicineMedical Imaging Techniques and ApplicationsAdvanced X-ray and CT ImagingDental Radiography and Imaging
Volumetric computed tomography with carbon nanotube X-ray source array for improved image quality and accuracy | Litcius