Improving Harmonization and Standardization of Expanded Newborn Screening Results by Optimization of the Legacy Flow Injection Analysis Tandem Mass Spectrometry Methods and Application of a Standardized Calibration Approach

Rachel S. Carling, Emily Whyte, Catharine John, Rachelle Garstone, Philippa Goddard, Toby Greenfield, S. Hogg, Clare Le Masurier, Simon Cowen, Stuart J. Moat, Christopher Hopley

2022Clinical Chemistry14 citationsDOIOpen Access PDF

Abstract

BACKGROUND: Newborn screening (NBS) laboratories in the United Kingdom adhere to common protocols based on single analyte cutoff values (COVs); therefore, interlaboratory harmonization is of paramount importance. Interlaboratory variation for screening analytes in UK NBS laboratories ranges from 17% to 59%. While using common stable isotope internal standards has been shown to significantly reduce interlaboratory variation, instrument set-up, sample extraction, and calibration approach are also key factors. METHODS: Dried blood spot (DBS) extraction processes, instrument set-up, mobile-phase composition, sample introduction technique, and calibration approach of flow injection analysis-tandem mass spectrometry (FIA-MS/MS) methods were optimized. Inter- and intralaboratory variation of methionine, leucine, phenylalanine, tyrosine, isovaleryl-carnitine, glutaryl-carnitine, octanoyl-carnitine, and decanoyl-carnitine were determined pre- and postoptimization, using 3 different calibration approaches. RESULTS: Optimal recovery of analytes from DBS was achieved with a 35-min extraction time and 80% methanol (150 μL). Optimized methodology decreased the mean intralaboratory percentage relative SD (%RSD) for the 8 analytes from 20.7% (range 4.1-46.0) to 5.4% (range 3.0-8.5). The alternative calibration approach reduced the mean interlaboratory %RSD for all analytes from 16.8% (range 4.1-25.0) to 7.1% (range 4.1-11.0). Nuclear magnetic resonance analysis of the calibration material highlighted the need for standardization. The purities of isovaleryl-carnitine and glutaryl-carnitine were 85.13% and 69.94% respectively, below the manufacturer's stated values of ≥98%. CONCLUSIONS: For NBS programs provided by multiple laboratories using single analyte COVs, harmonization and standardization of results can be achieved by optimizing legacy FIA-MS/MS methods, adopting a common analytical protocol, and using standardized calibration material rather than internal calibration.

Topics & Concepts

AnalyteChromatographyCalibrationCalibration curveCoefficient of variationChemistryMass spectrometrySample preparationTandem mass spectrometryExtraction (chemistry)Detection limitMathematicsStatisticsMetabolism and Genetic DisordersMetabolomics and Mass Spectrometry StudiesBiosimilars and Bioanalytical Methods