Through-Focus Response of Extended Depth of Focus Intraocular Lenses

Stefan Pieh, Cornelia Artmayr, Viktoria Pai, Daniel Schartmüller, Katharina Kriechbaum

2022Journal of Refractive Surgery19 citationsDOI

Abstract

PURPOSE: To investigate extended depth of focus intraocular lenses (EDOF IOLs), intended to extend the depth of field after cataract surgery in a comparative study at the optical bench. METHODS: An optical bench with either green or white light was used for this study. The artificial cornea used exhibited a spherical aberration of 0.2 µm. Examinations of the following lenses with a pupil opening of 3 and 4.5 mm were carried out: AcrySof IQ Vivity (Alcon Laboratories, Inc), Isopure (PhysIOL), Tecnis Eyhance (Johnson & Johnson), Vivinex Impress (Hoya Surgical Optics), and xact (Santen). RESULTS: Using green light and a pupil aperture of 3 mm, the AcrySof IQ Vivity showed the highest light energy for the intermediate area, whereas the Isopure and Vivinex Impress provided the highest energy for distance vision. Under the same examination conditions with a pupil opening of 4.5 mm, all lenses showed a low light distribution for the intermediate range. Regarding light distribution for distance, the Tecnis Eyhance had the highest light intensity. Using white light, the curves became much wider and more similar to each other. CONCLUSIONS: .