Sources of Inaccuracy in Boron Isotope Measurement Using <scp>LA‐MC‐ICP‐MS</scp>

Jan Fietzke, Eleni Anagnostou

2023Geostandards and Geoanalytical Research11 citationsDOIOpen Access PDF

Abstract

Laser ablation multi‐collector‐inductively coupled plasma‐mass spectrometry (LA‐MC‐ICP‐MS) has become a valuable tool for the in situ measurement of the boron isotope composition of geological samples at high (tens to hundreds of μm) spatial resolution. That said, this application suffers from significant analytical challenges. We focus in this study on the underlying processes of two of the main causes for inaccuracies using this technique. We provide empirical evidence that not only Ca ions (Sadekov et al . 2019, Standish et al . 2019, Evans et al . 2021) but also Ar ions, that are reflected within the flight tube of the mass spectrometer, are the source for previously reported issues with spectral baselines. We also address the impact of plasma conditions on the instrumental mass fractionation as a source for matrix‐ and mass‐load‐related analytical biases. Comparing experimental data with the results of a dedicated release and diffusion model (RDM) we estimate that a close to complete (~ 97%) release of boron from the sample aerosol is needed to allow for consistently accurate LA boron isotope measurement results without the need for corrections.

Topics & Concepts

IsotopeAnalytical Chemistry (journal)Mass spectrometryBoronChemistryIsotopes of boronLaser ablationInductively coupled plasma mass spectrometryIonFractionationLaserPhysicsOpticsEnvironmental chemistryChromatographyNuclear physicsOrganic chemistryRadioactive element chemistry and processingRadioactive contamination and transferIsotope Analysis in Ecology