Litcius/Paper detail

Long-Term Electrode–Skin Impedance Variation for Electromyographic Measurements

Andreia S. P. Sousa, Andreia Noites, Rui Vilarinho, Rubim Santos

2023Sensors11 citationsDOIOpen Access PDF

Abstract

This study aims to observe the evolution of the electrode–skin interface impedance of surface EMG electrodes over the time taken to determine the time of stabilization. Eight healthy subjects participated in the study. Electrode–skin impedance was evaluated in the rectus abdominal muscle every five minutes, over a total period of 50 min. A reduction of 13.23% in the impedance values was observed in minute 10 (p = 0.007), and a reduction of 9.02% was observed in minute 15 (p = 0.029). No statistically significant differences were observed in the other instants evaluated. The findings obtained in the present study demonstrate a decrease in electrode–skin impedance from minute 5 to minute 15, followed by a stabilization period with a low percentage of variation till minute 50.

Topics & Concepts

ElectrodeElectrical impedanceBiomedical engineeringFocused Impedance MeasurementMedicineMaterials scienceChemistryElectrical engineeringEngineeringPhysical chemistryMuscle activation and electromyography studiesNeuroscience and Neural EngineeringAdvanced Sensor and Energy Harvesting Materials