Selective Detection of Volatile Organics in a Mixture Using a Photoionization Detector and Thermal Desorption from a Nanoporous Preconcentrator
Joshua Prestage, Coco Day, Shamus L.G. Husheer, William T. Winter, Wah On Ho, John Saffell, Tanya Hutter
Abstract
The selective detection of individual hazardous volatile organic compounds (VOCs) within a mixture is of great importance in industrial contexts due to environmental and health concerns. Achieving this with inexpensive, portable detectors continues to be a significant challenge. Here, a novel thermal separator system coupled with a photoionization detector has been developed, and its ability to selectively detect the VOCs isopropanol and 1-octene from a mixture of the two has been studied. The system includes a nanoporous silica preconcentrator in conjunction with a commercially available photoionization detector (PID). The PID is a broadband total VOC sensor with little selectivity; however, when used in conjunction with our thermal desorption approach, selective VOC detection within a mixture can be achieved. VOCs are adsorbed in the nanoporous silica over a 5 min period at 5 °C before being desorbed by heating at a fixed rate to 70 °C and detected by the PID. Different VOCs desorb at different times/temperatures, and mathematical analysis of the set of PID responses over time enabled the contributions from isopropanol and 1-octene to be separated. The concentrations of each compound individually could be measured in a mixture with limits of detection less than 10 ppbv and linearity errors less than 1%. Demonstration of a separation of a mixture of chemically similar compounds, benzene and o-xylene, is also provided.