Cochlear Implantation in Infants: Evidence of Safety
Nicholas L. Deep, Patricia L. Purcell, Karen A. Gordon, Blake C. Papsin, J. Thomas Roland, Susan B. Waltzman
Abstract
The aim of this study was to investigate surgical, anesthetic, and device-related complications associated with cochlear implantation (CI) in children younger than 1 year of age. This was a multicenter, retrospective chart review of all children with severe-to-profound sensorineural hearing loss who underwent cochlear implantation with a Cochlear Nucleus Implant System before 1 year of age. Endpoints included perioperative course, major and minor surgical, anesthetic and device-related complications, and 30-day readmission rates. One hundred thirty-six infants (242 ears) met criteria. The mean age at implantation was 9.4 months (standard deviation 1.8). Six-month follow-up was reported in all patients. There were no major anesthetic or device-related complications. Adverse events were reported in 34 of implanted ears (14%; 7 major, 27 minor). Sixteen adverse events occurred ≤30 days of surgery, and 18 occurred >30 days of surgery. The 30-day readmission rate was 1.5%. The rate of adverse events did not correlate with preexisting medical comorbidities or duration under anesthesia. There was no significant difference detected in complication rate for patients younger than 9 months of age versus those 9 to 11 months of age. This study demonstrates the safety of CI surgery in infants and supports reducing the indication for cochlear implantation to younger than 1 year of age for children with bilateral, profound sensorineural hearing loss obtaining a Cochlear Nucleus Implant System.