Litcius/Paper detail

Performance of Spectral Photon-Counting Coronary CT Angiography and Comparison with Energy-Integrating-Detector CT: Objective Assessment with Model Observer

David C. Rotzinger, Damien Racine, Fabio Becce, Elias Lahoud, Klaus Erhard, Salim Si‐Mohamed, Joël Greffier, Anaïs Viry, Loïc Boussel, Reto Meuli, Yoad Yagil, Pascal Monnin, Philippe Douek

2021Diagnostics41 citationsDOIOpen Access PDF

Abstract

AIMS: To evaluate spectral photon-counting CT's (SPCCT) objective image quality characteristics in vitro, compared with standard-of-care energy-integrating-detector (EID) CT. METHODS: We scanned a thorax phantom with a coronary artery module at 10 mGy on a prototype SPCCT and a clinical dual-layer EID-CT under various conditions of simulated patient size (small, medium, and large). We used filtered back-projection with a soft-tissue kernel. We assessed noise and contrast-dependent spatial resolution with noise power spectra (NPS) and target transfer functions (TTF), respectively. Detectability indices (d') of simulated non-calcified and lipid-rich atherosclerotic plaques were computed using the non-pre-whitening with eye filter model observer. RESULTS: ). In the detectability analysis, SPCCT outperformed EID-CT in all investigated conditions, providing superior d'. SPCCT reached almost perfect detectability (AUC ≈ 95%) for simulated 0.5-mm-thick non-calcified plaques (for large-sized patients), whereas EID-CT had lower d' (AUC ≈ 75%). For lipid-rich atherosclerotic plaques, SPCCT achieved 85% AUC vs. 77.5% with EID-CT. CONCLUSIONS: SPCCT outperformed EID-CT in detecting simulated coronary atherosclerosis and might enhance diagnostic accuracy by providing lower noise magnitude, markedly improved spatial resolution, and superior lipid core detectability.

Topics & Concepts

Observer (physics)DetectorPhotonPhoton countingCoronary angiographyEnergy (signal processing)PhysicsAngiographyOpticsRadiologyMedicineNuclear medicineInternal medicineMyocardial infarctionQuantum mechanicsAdvanced X-ray and CT ImagingDigital Radiography and Breast ImagingAdvanced X-ray Imaging Techniques
Performance of Spectral Photon-Counting Coronary CT Angiography and Comparison with Energy-Integrating-Detector CT: Objective Assessment with Model Observer | Litcius