Litcius/Paper detail

Multicenter Evaluation of the QIAstat-Dx Respiratory Panel for Detection of Viruses and Bacteria in Nasopharyngeal Swab Specimens

Amy L. Leber, Jan Gorm Lisby, Glen Hansen, Ryan F. Relich, Uffe Vest Schneider, Paul A. Granato, Stephen Young, Josep Pareja, Irene Hannet

2020Journal of Clinical Microbiology58 citationsDOIOpen Access PDF

Abstract

This multicenter evaluation provides data obtained from 1,994 prospectively collected and 310 retrospectively collected (archived) NPS specimens with performance compared to that of the BioFire FilmArray Respiratory Panel, version 1.7. The overall percent agreement between QIAstat-Dx RP and the comparator testing was 99.5%. In the prospective cohort, the QIAstat-Dx RP demonstrated a positive percent agreement of 94.0% or greater for the detection of all but four analytes: coronaviruses 229E, NL63, and OC43 and rhinovirus/enterovirus. The test also demonstrated a negative percent agreement of ≥97.9% for all analytes. The QIAstat-Dx RP is a robust and accurate assay for rapid, comprehensive testing for respiratory pathogens.

Topics & Concepts

RhinovirusMycoplasma pneumoniaeVirologyHuman metapneumovirusEnterovirusMicrobiologyCoronavirusVirusInfluenza-like illnessRespiratory tract infectionsMedicineBiologyRespiratory systemPneumoniaCoronavirus disease 2019 (COVID-19)Internal medicineInfectious disease (medical specialty)DiseaseRespiratory viral infections researchInfluenza Virus Research StudiesViral gastroenteritis research and epidemiology