Abstract 3‐dimensional (3D) Fe−Co−LDH/MXene composite was synthesized by in‐situ synthesis and assembly of Fe−Co−LDH rod around MXene under hydrothermal condition. Due to the unique 3D configuration and good conductivity, the obtained Fe−Co−LDH/MXene modified glassy carbon electrode (Fe−Co−LDH/MXene/GCE) showed excellent electrochemical activity for As(III) detection. Via square‐wave anodic stripping voltammetry, the response current on Fe−Co−LDH/MXene/GCE had good linear relationship with As(III) concentrations (1∼1000 ppt) with superior sensitivity (0.22 μA ppt −1 cm −2 ) and low detection limit (0.9 ppt). The mechanism of As(III) adsorption was demonstrated. The electrode showed excellent anti‐interference ability. Real water sample analysis demonstrated the Fe−Co−LDH/MXene/GCE was deployable in aqua‐system.

Topics & Concepts

Detection limitElectrochemistryArsenicElectrodeStripping (fiber)AdsorptionMaterials scienceVoltammetryChemistryNuclear chemistryAnalytical Chemistry (journal)Inorganic chemistryChromatographyMetallurgyOrganic chemistryPhysical chemistryComposite materialMXene and MAX Phase MaterialsCarbon and Quantum Dots ApplicationsAdvanced biosensing and bioanalysis techniques