Litcius/Paper detail

Feasibility and safety of dual-console telesurgery with the KangDuo Surgical Robot-1500 System using fifth-generation and wired networks: an animal experiment and sea-spanning clinical study

Liqing XU, Cheng Shen, Xinfei Li, Fangzhou Zhao, Wei Huang, Kunlin Yang, Peng Zhang, Yicong DU, Shubo Fan, Liang Cui, Liqun Zhou, Xuesong Li

2024Minerva Urology and Nephrology12 citationsDOI

Abstract

BACKGROUND: To evaluate the feasibility and safety of dual-console telesurgery with the new KangDuo system in an animal experiment and clinical study. METHODS: Six canine models were performed radical prostatectomy with dual-console KanDuo surgical robot-1500 (KD-SR-1500-RARP). The perioperative outcomes, physical and mental workload of the surgeon were collected. Physical workload was evaluated with surface electromyography. Mental workload was evaluated with NASA-TLX. After conducting animal experiments to verify safety of dual-console KD-SR-1500-RARP, we conducted the clinical trial using 5G and wired networks. RESULTS: In the animal experiment, all surgeries were performed successfully. The operative time was 80.2±32.1 min. The docking time was 2.4±0.5 min. The console time was 49.7±25.3 min. There were no perioperative complications or equipment related adverse events. All dogs can micturate after catheter removal at one week postoperatively. The mental workload was at a low level (a scale ranging from 0 to 60), which scored 15.7±6.9. Among the eight recorded muscles, the fatigue degree of the right radial flexor and left biceps was the highest two (iEMG, resection, 299.8±344 uV, 109.9±16.9 uV; suture, 849.4±1252.5 uV, 423.1±621.3 uV, respectively). In the clinical study, the console time was 136 min. The mean latency time was ≤200 ms. The data pocket loss was <1%. The operation was successfully completed without malfunctions occurring throughout the entire process. CONCLUSIONS: Dual-console telesurgery with the KD-SR-1500 system was shown to be feasible and safe in radical prostatectomy using 5G and wired networks.

Topics & Concepts

Surgical robotDual (grammatical number)RobotComputer scienceHuman–computer interactionMedicineArtificial intelligenceArtLiteratureSurgical Simulation and TrainingIntraoperative Neuromonitoring and Anesthetic EffectsSoft Robotics and Applications