Comparative evaluation of a dual-target real-time RT-PCR assay for COVID-19 diagnosis and assessment of performance in pooled saliva and nasopharyngeal swab samples
Cyril Chik‐Yan Yip, Kit‐Hang Leung, Anthony Chin‐Ki Ng, Kwok‐Hung Chan, Kelvin Kai‐Wang To, Jasper Fuk‐Woo Chan, Ivan Fan‐Ngai Hung, Vincent Chi‐Chung Cheng, Siddharth Sridhar
Abstract
OBJECTIVES: Sensitive molecular diagnostic assays are essential for COVID-19 diagnosis. We evaluated the Hecin Scientific SARS-CoV-2 nucleic acid test kit, a dual-target real-time RT-PCR assay targeting the SARS-CoV-2 N and ORF1ab genes. METHODS: The Hecin test kit's diagnostic performance in detecting SARS-CoV-2 RNA was compared to the LightMix Modular SARS and Wuhan CoV E-gene kit (TIB Molbiol) and an in-house single-tube nested real-time RT-PCR using 296 clinical specimens, 11 proficiency testing samples, and 30 low-positive deep throat saliva and nasopharyngeal swab (NPS) samples pooled into negative samples in ratios of 1:5, 1:10, and 1:30. RESULTS: The limit-of-detection of the Hecin test kit was around 500 dC/mL for the N and ORF1ab targets. Sensitivity and specificity of the Hecin test kit were 98.1% (95% CI: 93.4-99.8%) and 100% (98.1-100%), respectively, when measured against the reference method. The Hecin test kit showed fair sensitivity (80%) in low-positive NPS samples pooled in ratios of 1:5 and 1:10. Its performance in pooled samples could be dramatically improved by adjusting the assay Ct cutoff. CONCLUSION: The Hecin test kit enables sensitive and specific detection of SARS-CoV-2 in clinical samples and pooled samples.