Colorimetric sweat analysis using wearable hydrogel patch sensors for detection of chloride and glucose
Tuqiang Li, Xiaofeng Chen, Ying Fu, Caizhi Liao
Abstract
, and the glucose colorimetric sensor, with a detection limit down to 1 μM, exhibits an adequate linear response from 10 μM to 1 mM glucose. These colorimetric hydrogel sensors are also incorporated into a medical dressing to create wearable sensor devices for real-time sweat analysis. The acquired readings closely match the results obtained from the benchmark analyzing instrument, with a small deviation of less than 10%. Therefore, our simple colorimetric hydrogel sensing patches hold great potential to advance real-time sweat testing and contribute to the transitional development of wearable medical devices.
Topics & Concepts
SWEATWearable computerChemistryChlorideChromatographyComputer scienceOrganic chemistryInternal medicineEmbedded systemMedicineAdvanced Sensor and Energy Harvesting MaterialsAdvanced Chemical Sensor TechnologiesAnalytical Chemistry and Sensors