Litcius/Paper detail

105° field of view non-contact handheld swept-source optical coherence tomography

Shuibin Ni, Thanh-Tin P. Nguyen, Ringo Ng, Shanjida Khan, Susan Ostmo, Yali Jia, Michael F. Chiang, David Huang, J. Peter Campbell, Yifan Jian

2021Optics Letters29 citationsDOIOpen Access PDF

Abstract

We demonstrate a handheld swept-source optical coherence tomography (OCT) system with a 400 kHz vertical-cavity surface-emitting laser (VCSEL) light source, a non-contact approach, and an unprecedented single shot 105° field of view (FOV). We also implemented a spiral scanning pattern allowing real-time visualization with improved scanning efficiency. To the best of our knowledge, this is the widest FOV achieved in a portable non-contact OCT retinal imaging system to date. Improvements to the FOV may aid the evaluation of retinal diseases such as retinopathy of prematurity, where important vitreoretinal changes often occur in the peripheral retina.

Topics & Concepts

Optical coherence tomographyOpticsField of viewCoherence (philosophical gambling strategy)LaserComputer sciencePhysicsQuantum mechanicsOptical Coherence Tomography ApplicationsRetinal and Macular SurgeryRetinopathy of Prematurity Studies