Aktiia cuffless blood pressure monitor yields equivalent daytime blood pressure measurements compared to a 24-h ambulatory blood pressure monitor: Preliminary results from a prospective single-center study
Tiago P. Almeida, Meritxell Cortés, David Perruchoud, Jérémy Alexandre, Pascale Vermare, Josep Solà, Jay Shah, Luísa Marques, Cyril Pellaton
Abstract
Abstract In this preliminary study, we compared daytime blood pressure (BP) measurements performed by a commercially available cuffless—and continual—BP monitor (Aktiia monitor, Neuchâtel, Switzerland) and a traditional ambulatory BP monitor (ABPM; Dyasis 3, Novacor, Paris, France) from 52 patients enrolled in a 12-week cardiac rehabilitation (CR) program (Neuchâtel, Switzerland). Daytime (9am–9pm) systolic (SBP) and diastolic (DBP) BP from 7-day averaged data from Aktiia monitor were compared to 1-day averaged BP data from ABPM. No significant differences were found between the Aktiia monitor and the ABPM for SBP (μ ± σ [95% confidence interval]: 1.6 ± 10.5 [−1.5, 4.6] mmHg, P = 0.306; correlation [ R 2 ]: 0.70; ± 10/ ± 15 mmHg agreements: 60%, 84%). Marginally non-significant bias was found for DBP (−2.2 ± 8.0 [−4.5, 0.1] mmHg, P = 0.058; R 2 : 0.66; ±10/±15 mmHg agreements: 78%, 96%). These intermediate results show that daytime BP measurements using the Aktiia monitor generate data comparable to that of an ABPM monitor.