Aktiia cuffless blood pressure monitor yields equivalent daytime blood pressure measurements compared to a 24-h ambulatory blood pressure monitor: Preliminary results from a prospective single-center study

Tiago P. Almeida, Meritxell Cortés, David Perruchoud, Jérémy Alexandre, Pascale Vermare, Josep Solà, Jay Shah, Luísa Marques, Cyril Pellaton

2023Hypertension Research43 citationsDOIOpen Access PDF

Abstract

Abstract In this preliminary study, we compared daytime blood pressure (BP) measurements performed by a commercially available cuffless—and continual—BP monitor (Aktiia monitor, Neuchâtel, Switzerland) and a traditional ambulatory BP monitor (ABPM; Dyasis 3, Novacor, Paris, France) from 52 patients enrolled in a 12-week cardiac rehabilitation (CR) program (Neuchâtel, Switzerland). Daytime (9am–9pm) systolic (SBP) and diastolic (DBP) BP from 7-day averaged data from Aktiia monitor were compared to 1-day averaged BP data from ABPM. No significant differences were found between the Aktiia monitor and the ABPM for SBP (μ ± σ [95% confidence interval]: 1.6 ± 10.5 [−1.5, 4.6] mmHg, P = 0.306; correlation [ R 2 ]: 0.70; ± 10/ ± 15 mmHg agreements: 60%, 84%). Marginally non-significant bias was found for DBP (−2.2 ± 8.0 [−4.5, 0.1] mmHg, P = 0.058; R 2 : 0.66; ±10/±15 mmHg agreements: 78%, 96%). These intermediate results show that daytime BP measurements using the Aktiia monitor generate data comparable to that of an ABPM monitor.

Topics & Concepts

Ambulatory blood pressureBlood pressureMedicineAmbulatoryDaytimeCardiologyInternal medicineConfidence intervalDiastoleLimits of agreementProspective cohort studyNuclear medicineAtmospheric sciencesGeologyBlood Pressure and Hypertension StudiesHeart Rate Variability and Autonomic ControlCardiovascular Syncope and Autonomic Disorders
Aktiia cuffless blood pressure monitor yields equivalent daytime blood pressure measurements compared to a 24-h ambulatory blood pressure monitor: Preliminary results from a prospective single-center study | Litcius