Litcius/Paper detail

ST Genesia reference values of 117 healthy donors measured with STG‐BleedScreen, STG‐DrugScreen and STG‐ThromboScreen reagents

Marisa Ninivaggi, Romy M. W. de Laat-Kremers, Audrey Carlo, Bas de Laat

2020Research and Practice in Thrombosis and Haemostasis29 citationsDOIOpen Access PDF

Abstract

INTRODUCTION: The ST Genesia is a benchtop, fully automated thrombin generation (TG) device. It is completely standardized and ensures a uniform heat distribution throughout the measurement. We aimed to determine reference values and to compare TG in men and women with and without the use of oral contraceptives (OCs). MATERIALS AND METHODS: Plasma from 117 healthy donors was measured on the ST Genesia with the available reagent kits: STG-BleedScreen, STG-DrugScreen, and STG-ThromboScreen. All kits include at least two quality controls and a reference plasma to normalize data. STG-ThromboScreen has a second trigger containing thrombomodulin (TM) to include the effect on the protein C pathway. Means were compared with one-way analysis of variance and reference ranges were established with 2.5th to 97.5th percentiles on absolute TG parameters. RESULTS: < .05 compared to men/women without OCs). CONCLUSIONS: TG in men and women without OCs was similar; however, women taking OCs had significantly higher TG values, and the effect of TM was also less pronounced in these women.

Topics & Concepts

Library scienceCitationSynapseNeurosciencePsychologyComputer scienceBlood Coagulation and Thrombosis MechanismsReproductive Health and ContraceptionMaternal and fetal healthcare