The optimal monoenergetic spectral image level of coronary computed tomography (CT) angiography on a dual-layer spectral detector CT with half-dose contrast media

Xin Huang, Sizhe Gao, Yue Ma, Xiaomei Lü, Zheng Jia, Yang Hou

2020Quantitative Imaging in Medicine and Surgery36 citationsDOIOpen Access PDF

Abstract

BACKGROUND: To investigate the optimal monoenergetic level of spectral reconstructions in coronary computed tomography angiography (coronary CTA) on a dual-layer spectral detector computed tomography (SDCT) with half-dose contrast media. METHODS: Two hundred patients with suspected coronary artery disease (CAD) were enrolled in this prospective coronary CTA study and randomly divided into a routine-dose contrast media group and a half-dose contrast media group (each n=100). Coronary CTA was performed using SDCT with prospective electrocardiogram (ECG)-gated mode. A tube voltage of 120 kVp was used, along with an automated tube current modulation. A dose of iodixanol 270 mgI/mL of 0.8 and 0.4 mL/kg was administered to the routine and half-dose groups, respectively. For the routine-dose group, 120 kVp polychromatic images with a model-based iterative reconstruction (IMR) (Group A) were reconstructed. For the half-dose group, three monoenergetic levels of images were reconstructed (Group B, 45 keV; Group C, 50 keV; and Group D, 55 keV). Objective indicators [mean CT values; noise; signal-to-noise ratio (SNR); and contrast-to-noise ratio (CNR)] and subjective indicators (contrast, sharpness, subjective noise, and acceptability) in each group were compared. RESULTS: . 1.80±0.53 mSv, P=0.78) between the routine- and half-dose groups. The average iodine loads were 15.33±2.26 and 7.48±1.14 g, respectively. Mean CT values, SNR, CNR, and subjective contrast in Group C were higher than those in Group A (P<0.05), and there were no significant differences in other indicators between Group C and Group A (P>0.05). The objective and subjective noise in Group B were worse than those in Group A (P<0.05). The contrast, sharpness, and acceptability of Group D were all worse than those of Group A (P<0.05). CONCLUSIONS: Compared to routine polychromatic images, 50 keV monoenergetic images can provide equivalent or improved coronary image quality in coronary CTA performed on SDCT with half the amount of contrast media.

Topics & Concepts

MedicineIodixanolNuclear medicineCoronary artery diseaseImage noiseTomographyEffective dose (radiation)Contrast-to-noise ratioContrast (vision)Computed tomography angiographyRadiation doseAngiographyContrast mediumRadiologyImage qualityPhysicsOpticsInternal medicineComputer scienceImage (mathematics)Artificial intelligenceAdvanced X-ray and CT ImagingCardiac Imaging and DiagnosticsDigital Radiography and Breast Imaging