Radiofrequency echographic multi-spectrometry and DXA for the evaluation of bone mineral density in a peritoneal dialysis setting
Angelo Fassio, Stefano Andreola, Davide Gatti, Beatrice Bianco, Matteo Gatti, Giovanni Gambaro, Maurizio Rossini, Ombretta Viapiana, Riccardo Negrelli, Giovanni Adami
Abstract
BACKGROUND: The aim of this real-life cross-sectional explorative study was to compare radiofrequency echographic multi-spectrometry (REMS) with dual-energy X-rays absorptiometry (DXA) in the BMD assessment of patients receiving peritoneal dialysis (PD). Furthermore, we investigated the relationship between lumbar aortic calcifications (AOCs) and the DXA lumbar measurements. METHODS: ) were compared. Cohen's k coefficients were used to assess the inter-technique agreement in the classification of patients as osteoporotic. Lumbar AOCs were estimated semi-quantitatively on laterolateral DXA scans. RESULTS: 41 patients were enrolled. No significant differences were documented between the BMD T-scores measured through DXA or REMS at the femur. At the lumbar spine, the DXA anteroposterior mean T-score (- 0.49 ± 1.98) was significantly higher than both the laterolateral DXA (- 1.66 ± 0.99) and the REMS (- 2.00 ± 1.94) measurements (p < 0.01 vs both). No significant differences were found between the DXA and REMS fracture risk estimates with both algorithms. The inter-technique Cohen's k coefficient (for the worst T-score, any site) was 0.421, p < 0.001. The discrepancy between the DXA laterolateral and anteroposterior lumbar T-score was positively associated with the AOCs extent and severity (r = 0.402, p < 0.01). CONCLUSIONS: Our data showed a promising agreement, in a real-life PD setting, between DXA and REMS BMD assessment and in the consequent fracture risk estimation and confirm the AOCs interference on the diagnostic accuracy of lumbar DXA.