Abstract Phosphocreatine (PCr) plays a vital role in neuron and myocyte energy homeostasis. Currently, there are no routine diagnostic tests to noninvasively map PCr distribution with clinically relevant spatial resolution and scan time. Here, we demonstrate that artificial neural network-based chemical exchange saturation transfer (ANNCEST) can be used to rapidly quantify PCr concentration with robust immunity to commonly seen MRI interferences. High-quality PCr mapping of human skeletal muscle, as well as the information of exchange rate, magnetic field and radio-frequency transmission inhomogeneities, can be obtained within 1.5 min on a 3 T standard MRI scanner using ANNCEST. For further validation, we apply ANNCEST to measure the PCr concentrations in exercised skeletal muscle. The ANNCEST outcomes strongly correlate with those from 31 P magnetic resonance spectroscopy ( R = 0.813, p < 0.001, t test). These results suggest that ANNCEST has potential as a cost-effective and widely available method for measuring PCr and diagnosing related diseases.

Topics & Concepts

PhosphocreatineMagnetic resonance imagingNuclear magnetic resonanceSkeletal muscleBiomedical engineeringBiologyMedicineEnergy metabolismInternal medicinePhysicsRadiologyAdvanced MRI Techniques and ApplicationsLanthanide and Transition Metal ComplexesAtomic and Subatomic Physics Research