We evaluated the performance of the interoperable Cambridge hybrid closed-loop app with FreeStyle Libre 3 glucose sensor, and YpsoPump insulin pump in a real-world setting. Data from 100 users (63 adults [mean ± SD age 41.9 ± 14.0 years], 15 children [8.6 ± 5.2 years)] and 22 users of unreported age) for a period of 28 days were analyzed. Time in range (3.91- 10.0mmol/L) was 72.6 ± 11.1% overall. Time below range (<3.9mmol/L) was 3.1% (1.4-5.1) (median [interquartile range]). Auto-mode was active for 95.8% (91.8-97.9) of time. This real-world analysis suggests that the performance of Cambridge hybrid closed-loop app with this glucose sensor is comparable to other commercially available hybrid closed-loop systems.

Topics & Concepts

Interquartile rangeClosed loopContinuous glucose monitoringInteroperabilityMedicineComputer scienceDiabetes mellitusInternal medicineEndocrinologyOperating systemType 1 diabetesEngineeringControl engineeringDiabetes Management and ResearchWireless Body Area NetworksMobile Health and mHealth Applications