Measurement of therapeutic <sup>12</sup> C beam in a water phantom using CR-39

Shiyan Yang, Jingfang Zhao, Weihai Zhuo, H. F. Shen, Bo Chen

2021Journal of Radiological Protection16 citationsDOI

Abstract

mono-energy carbon beam at the heavy ion facility in the Shanghai Proton and Heavy Ion Center. The spatial distribution of the primary carbon beam and secondary fragments in a water phantom were systematically analyzed both in the transverse direction (perpendicular to the projection direction of the primary beam) and at different depths in the longitudinal direction (along the projection direction of the primary beam) with measured tracks on the CR-39 detectors. Meanwhile, the theoretically spatial distribution and linear energy transfer (LET) spectra of the primary beam and secondary fragments were calculated using the Monte Carlo (MC) toolkit Geant4. The results showed that the CR-39 detectors are capable of providing high lateral resolution of carbon ion at different depths. In the range of the primary carbon beam, the beam width simulated with MC is in good agreement with that of experimental measurement. The track size registered in the CR-39 has a good correlation with the particle LET. These findings indicate that the CR-39 can be used for measuring both the particle flux and its spatial distribution of carbon ions.

Topics & Concepts

Beam (structure)Linear energy transferProtonImaging phantomMonte Carlo methodIonCarbon fibersMaterials scienceDetectorIrradiationIon beamPhysicsOpticsAtomic physicsNuclear physicsStatisticsComposite materialMathematicsQuantum mechanicsComposite numberRadiation Therapy and DosimetryRadiation Detection and Scintillator TechnologiesRadiation Effects in Electronics