Is There Still a Role for Two-Phase Contrast-Enhanced CT and Virtual Monoenergetic Images in the Era of Photon-Counting Detector CT?

Arne Estler, Konstantin Nikolaou, Stefan O. Schönberg, Fabian Bamberg, Matthias F. Froelich, Fabian Tollens, Niklas Verloh, Jakob Weiß, Marius Horger, Florian Hagen

2023Diagnostics12 citationsDOIOpen Access PDF

Abstract

Background: To compare the diagnostic characteristics between arterial phase imaging versus portal venous phase imaging, applying polychromatic T3D images and low keV virtual monochromatic images using a 1st generation photon-counting CT detector, of CT in patients with hepatocellular carcinoma (HCC). Methods: Consecutive patients with HCC, with a clinical indication for CT imaging, were prospectively enrolled. Virtual monoenergetic images (VMI) were reconstructed at 40 to 70 keV for the PCD-CT. Two independent, blinded radiologists counted all hepatic lesions and quantified their size. The lesion-to-background ratio was quantified for both phases. SNR and CNR were determined for T3D and low VMI images; non-parametric statistics were used. Results: Among 49 oncologic patients (mean age 66.9 ± 11.2 years, eight females), HCC was detected in both arterial and portal venous scans. The signal-to-noise ratio, the CNR liver-to-muscle, the CNR tumor-to-liver, and CNR tumor-to-muscle were 6.58 ± 2.86, 1.40 ± 0.42, 1.13 ± 0.49, and 1.53 ± 0.76 in the arterial phase and 5.93 ± 2.97, 1.73 ± 0.38, 0.79 ± 0.30, and 1.36 ± 0.60 in the portal venous phase with PCD-CT, respectively. There was no significant difference in SNR between the arterial and portal venous phases, including between “T3D” and low keV images (p > 0.05). CNRtumor-to-liver differed significantly between arterial and portal venous contrast phases (p < 0.005) for both “T3D” and all reconstructed keV levels. CNRliver-to-muscle and CNRtumor-to-muscle did not differ in either the arterial or portal venous contrast phases. CNRtumor-to-liver increased in the arterial contrast phase with lower keV in addition to SD. In the portal venous contrast phase, CNRtumor-to-liver decreased with lower keV; whereas, CNRtumor-to-muscle increased with lower keV in both arterial and portal venous contrast phases. CTDI and DLP mean values for the arterial upper abdomen phase were 9.03 ± 3.59 and 275 ± 133, respectively. CTDI and DLP mean values for the abdominal portal venous phase were 8.75 ± 2.99 and 448 ± 157 with PCD-CT, respectively. No statistically significant differences were found concerning the inter-reader agreement for any of the (calculated) keV levels in either the arterial or portal-venous contrast phases. Conclusions: The arterial contrast phase imaging provides higher lesion-to-background ratios of HCC lesions using a PCD-CT; especially, at 40 keV. However, the difference was not subjectively perceived as significant.

Topics & Concepts

Hepatocellular carcinomaMedicineNuclear medicineContrast-to-noise ratioRadiologyMonochromatic colorImage qualityPhysicsInternal medicineOpticsImage (mathematics)Artificial intelligenceComputer scienceAdvanced X-ray and CT ImagingAdvanced X-ray Imaging TechniquesRadiation Shielding Materials Analysis
Is There Still a Role for Two-Phase Contrast-Enhanced CT and Virtual Monoenergetic Images in the Era of Photon-Counting Detector CT? | Litcius