Clinical Validation of Automated Corrected QT-Interval Measurements From a Single Lead Electrocardiogram Using a Novel Smartwatch

Diego Mannhart, Elisa Hennings, Mirko Lischer, Claudius Vernier, Jeanne du Fay de Lavallaz, Sven Knecht, Beat Schaer, Stefan Osswald, Michael Kühne, Christian Sticherling, Patrick Badertscher

2022Frontiers in Cardiovascular Medicine16 citationsDOIOpen Access PDF

Abstract

Introduction: The Withings Scanwatch (Withings SA, Issy les Moulineaux, France) offers automated analysis of the QTc. We aimed to compare automated QTc-measurements using a single lead ECG of a novel smartwatch (Withings Scanwatch, SW-ECG) with manual-measured QTc from a nearly simultaneously recorded 12-lead ECG. Methods: We enrolled consecutive patients referred to a tertiary hospital for cardiac workup in a prospective, observational study. The QT-interval of the 12-lead ECG was manually interpreted by two blinded, independent cardiologists through the tangent-method. Bazett's formula was used to calculate QTc. Results were compared using the Bland-Altman method. Results: A total of 317 patients (48% female, mean age 63 ± 17 years) were enrolled. HR-, QRS-, and QT-intervals were automatically calculated by the SW in 295 (93%), 249 (79%), and 177 patients (56%), respectively. Diagnostic accuracy of SW-ECG for detection of QTc-intervals ≥ 460 ms (women) and ≥ 440 ms (men) as quantified by the area under the curve was 0.91 and 0.89. The Bland-Altman analysis resulted in a bias of 6.6 ms [95% limit of agreement (LoA) -59 to 72 ms] comparing automated QTc-measurements (SW-ECG) with manual QTc-measurement (12-lead ECG). In 12 patients (6.9%) the difference between the two measurements was greater than the LoA. Conclusion: In this clinical validation of a direct-to-consumer smartwatch we found fair to good agreement between automated-SW-ECG QTc-measurements and manual 12-lead-QTc measurements. The SW-ECG was able to automatically calculate QTc-intervals in one half of all assessed patients. Our work shows, that the automated algorithm of the SW-ECG needs improvement to be useful in a clinical setting.

Topics & Concepts

QT intervalMedicineQRS complexLimits of agreementInternal medicineElectrocardiographyCardiologySmartwatchNuclear medicineComputer scienceWearable computerEmbedded systemECG Monitoring and AnalysisCardiac electrophysiology and arrhythmiasAcute Myocardial Infarction Research