Disappointing long-term outcome of THA with carbon-fiber-reinforced poly-ether-ether-ketone (CFR-PEEK) as acetabular insert liner: a prospective study with a mean follow-up of 14.3 years
Luc J.M. Heijnens, Martijn G.M. Schotanus, A. D. Verburg, E. H. van Haaren
Abstract
INTRODUCTION: Insert liner wear of the acetabular component is one of the predictive values for survival of total hip arthroplasties (THAs). This prospective single-centre study was designed to evaluate the follow-up of carbon-fiber-reinforced poly-ether-ether-ketone (CFR-PEEK) insert liner used as bearing in cementless THAs. METHODS: 29 healthy patients with an indication for cementless THA were selected for a CFR-PEEK insert liner and followed over time. All patients received a cementless THA with a CFR-PEEK insert liner used as bearing. At different follow-up moments patients were routinely examined and were analysed using the Oxford Hip Score (OHS), the modified Merle d'Aubigne-Postel (MAP) score, and radiologically. At the follow up moments the plain radiographics where assessed for loosening, cyst formations and wear of the CFR-PEEK liners. RESULTS: At a mean of 14.3 years follow-up 4 revisions of the acetabular component were performed, resulting in a survival rate of 86.5% (CI 95%, 72.4-96.6). A statistically significant difference in OHS and MAP scores between pre- and postoperative follow-up moments was observed. The acetabular components of the remaining patients showed no radiological abnormalities at 14.3 years follow-up. The overall CFR-PEEK wear was low, with a mean of 0.81 (0.2-1.4) mm wear at 14.3 years follow-up. CONCLUSIONS: without abnormalities at 14.3 years follow-up.