Reliability of resting metabolic rate between and within day measurements using the Vyntus CPX system and comparison against predictive formulas

Juma Iraki, Gøran Paulsen, Ina Garthe, Gary Slater, José L. Areta

2021Nutrition and Health11 citationsDOIOpen Access PDF

Abstract

Background: To detect longitudinal changes of resting metabolic rate (RMR) resulting from the effects of energetic stress, reliable RMR measurements are crucial. The Vyntus CPX is a new automated indirect calorimetry system for which RMR reliability has not been determined. Additionally, its agreement with common predictive RMR formulas is unknown. Aim: To determine the within and between-day reliability of RMR measurements using the Vyntus CPX system and its agreement with predictive RMR formulas. Methods: Young (31 ± 7 years) healthy participants (n = 26, 12 females, 14 males) completed three measurements of RMR, two consecutive measures on the same day, one the day before/after, all under standardised conditions. Reliability was assessed with pairwise comparisons of between-day at the same time (BDST), within day consecutive measurements (WDCM) and between-day different time (BDDT), for parameters of reliability (mean change (MC), intraclass correlation (ICC) and typical error of measurement (TEM)). Measured RMR values (kcal/day) were compared against predictive values of 4 common formulas. Results: Parameters of reliability (mean, (95% confidence interval)) were: -BDST: MC, 0.2(-2.3—2.7)% (p = 0.67); ICC, 0.92(0.84—0.97); TEM, 4.5(3.5—6.2)%. -WDCM: MC, −2.5(-6.2—1.3)% (p = 0.21); ICC, 0.88(0.74—0.88); TEM, 7.0(5.4—9.8)%. -BDDT: MC, −1.5(-4.8—1.9)% (p = 0.57); ICC, 0.90(0.76—0.95); TEM, 6.1(4.8—8.5)%. RMR ratios (measured/predicted) were: 1.04 ± 0.14 (Nelson, p = 0.13), 1.03 ± 0.10 (Mifflin, p = 0.21), 0.98 ± 0.09 (Harris-benedict, p = 0.30), 0.95 ± 0.11 (Cunningham 1980 , p = 0.01), 1.00 ± 0.12 (Cunningham 1991 , p = 0.90) and 0.96 ± 0.13 (DXA, p = 0.03). Conclusions: The Vyntus CPX is reliable and measured RMR values agreed with four predictive formulas but are lower than Cunningham 1980 and DXA RMR estimates for this population.

Topics & Concepts

Reliability (semiconductor)Reliability engineeringStatisticsMedicineMathematicsEngineeringPhysicsQuantum mechanicsPower (physics)Clinical Nutrition and GastroenterologyDiet and metabolism studiesMuscle metabolism and nutrition
Reliability of resting metabolic rate between and within day measurements using the Vyntus CPX system and comparison against predictive formulas | Litcius