Low tibial baseplate migration 1 year after unrestricted kinematically aligned total knee arthroplasty using a medial conforming implant design
Abigail E. Niesen, Anna L. Garverick, Stephen M. Howell, Maury L. Hull
Abstract
PURPOSE: Varus alignment of the tibial baseplate and limb > 3° might adversely affect baseplate fixation after total knee arthroplasty (TKA), especially for unrestricted kinematically aligned (KA) TKA which aligns a majority of baseplates in varus. The purposes of this study were to determine whether baseplate migration at 1 year (1) was significantly less than a stability limit of 0.5 mm, (2) increased over time, and (3) was related to varus alignment of the baseplate and limb after unrestricted KA TKA. METHODS: Thirty-five patients underwent unrestricted KA TKA using a fixed-bearing, cemented, medial conforming tibial insert with posterior cruciate ligament retention. Using model-based radiostereometric analysis, maximum total point motion (MTPM) (i.e., largest displacement on the baseplate) was computed at 6 weeks, 3 months, 6 months, and 1 year postoperatively relative to the day of surgery. Baseplate and limb alignment were measured postoperatively on long-leg CT scanograms. RESULTS: At 1 year, mean MTPM of 0.35 mm was significantly less than the 0.5 mm stability limit (p = 0.0002). Mean MTPM did not increase from 6 weeks to 1 year (p = 0.3047). Notably, 89% (31/35) of tibial baseplates and 46% (16/35) of limbs were > 3° varus. Baseplate and limb alignment had no relationship to MTPM at 1 year (|r|≤ 0.173, p ≥ 0.3276). CONCLUSION: Low and non-progressive tibial baseplate migration 1 year after unrestricted KA TKA with a medial conforming design should allay any concern that unrestricted KA TKA increases risk of baseplate loosening due to varus alignment of the baseplate and limb. LEVEL OF EVIDENCE: Level II, therapeutic prospective cohort study.