Detection of Atrial Fibrillation Using a Ring-Type Wearable Device (CardioTracker) and Deep Learning Analysis of Photoplethysmography Signals: Prospective Observational Proof-of-Concept Study

Soonil Kwon, Joonki Hong, Eue‐Keun Choi, Byunghwan Lee, Changhyun Baik, Euijae Lee, Eui–Rim Jeong, Bon‐Kwon Koo, Seil Oh, Yung Yi

2020Journal of Medical Internet Research66 citationsDOIOpen Access PDF

Abstract

BACKGROUND: Continuous photoplethysmography (PPG) monitoring with a wearable device may aid the early detection of atrial fibrillation (AF). OBJECTIVE: We aimed to evaluate the diagnostic performance of a ring-type wearable device (CardioTracker, CART), which can detect AF using deep learning analysis of PPG signals. METHODS: Patients with persistent AF who underwent cardioversion were recruited prospectively. We recorded PPG signals at the finger with CART and a conventional pulse oximeter before and after cardioversion over a period of 15 min (each instrument). Cardiologists validated the PPG rhythms with simultaneous single-lead electrocardiography. The PPG data were transmitted to a smartphone wirelessly and analyzed with a deep learning algorithm. We also validated the deep learning algorithm in 20 healthy subjects with sinus rhythm (SR). RESULTS: In 100 study participants, CART generated a total of 13,038 30-s PPG samples (5850 for SR and 7188 for AF). Using the deep learning algorithm, the diagnostic accuracy, sensitivity, specificity, positive-predictive value, and negative-predictive value were 96.9%, 99.0%, 94.3%, 95.6%, and 98.7%, respectively. Although the diagnostic accuracy decreased with shorter sample lengths, the accuracy was maintained at 94.7% with 10-s measurements. For SR, the specificity decreased with higher variability of peak-to-peak intervals. However, for AF, CART maintained consistent sensitivity regardless of variability. Pulse rates had a lower impact on sensitivity than on specificity. The performance of CART was comparable to that of the conventional device when using a proper threshold. External validation showed that 94.99% (16,529/17,400) of the PPG samples from the control group were correctly identified with SR. CONCLUSIONS: A ring-type wearable device with deep learning analysis of PPG signals could accurately diagnose AF without relying on electrocardiography. With this device, continuous monitoring for AF may be promising in high-risk populations. TRIAL REGISTRATION: ClinicalTrials.gov NCT04023188; https://clinicaltrials.gov/ct2/show/NCT04023188.

Topics & Concepts

PhotoplethysmogramAtrial fibrillationMedicineCartWearable computerSinus rhythmCardioversionCardiologyElectrocardiographyInternal medicineWearable technologyProspective cohort studyPulse (music)Computer scienceComputer visionMechanical engineeringDetectorTelecommunicationsEmbedded systemFilter (signal processing)EngineeringNon-Invasive Vital Sign MonitoringAtrial Fibrillation Management and OutcomesECG Monitoring and Analysis