Analytical validation of white blood cell differential and platelet assessment on the Sysmex <scp>DI</scp>‐60 digital morphology analyzer

Ivana Lapić, Marija Miloš, Marija Dorotić, Valentina Drenški, Désirée Coen Herak, Dunja Rogić

2023International Journal of Laboratory Hematology11 citationsDOI

Abstract

INTRODUCTION: Digital morphology analyzers are increasingly replacing light microscopy in laboratory hematology practice. This study aimed to perform the analytical validation of the white blood cell (WBC) differential and of reliability of platelet assessment on Sysmex DI-60 (Kobe, Japan). METHODS: Validation included determination of within-run and between-run precision for WBC differential according to the CLSI EP15-A3 protocol, accuracy and method comparison with light microscopy and with the automated WBC differential from the Sysmex XN-10 hematology analyzer, reliability of platelet clump detection and platelet count estimation. RESULTS: Standard deviations of both pre- and post-classification mostly satisfied manufacturer's criteria for imprecision. Accuracy assessment revealed that only eosinophil count (1.4%) in one peripheral blood smear (PBS) remained outside the declared range (2-10%) after reclassification. Method comparison between DI-60 and light microscopy yielded Spearman's correlation coefficients from 0.37 (basophils) to 0.94 (neutrophils and lymphocytes), minor proportional difference for bands, constant difference for monocytes, both constant and proportional difference for lymphocytes and statistically significant biases for bands, lymphocytes, monocytes and basophils. Diagnostic sensitivity (Se) and specificity (Sp) of DI-60 in detecting immature/pathological cells were 88.7% (95%CI:81.1-94.0) and 83.0% (95%CI:78.7-86.7), respectively, with the area under the curve (AUC) of 0.86 (95%CI:0.82-0.89). Agreement in detection of platelet clumps was 94.8% (kappa coefficient = 0.67, 95%CI:0.53-0.80). Se and Sp of DI-60 to detect platelet clumps were 65.7% (95%CI: 47.8-80.9) and 96.9% (95%CI: 93.9-98.6), respectively, while AUC was 0.81 (95%CI: 0.76-0.86). CONCLUSION: DI-60 provides reliable WBC differential and platelet assessment. In doubtful cases, the use of light microscopy is still mandatory.

Topics & Concepts

Hematology analyzerWhite blood cellPlateletMedicineKappaNuclear medicineSignificant differenceEosinophilCoefficient of variationPathologyHematologyInternal medicineCell countingArea under the curvePeripheral bloodGastroenterologyMathematicsChromatographyChemistryGeometryCell cycleCancerAsthmaClinical Laboratory Practices and Quality ControlDigital Imaging for Blood DiseasesHematological disorders and diagnostics
Analytical validation of white blood cell differential and platelet assessment on the Sysmex <scp>DI</scp>‐60 digital morphology analyzer | Litcius